7550 Precision Current Shunt
|
|
Lab Current Standard
1uA -250A Current Range AC / DC Current Shunt AC / DC Current Meter |
| Features | Applications |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Descriptions | |
|
Model
7550 has five precision selectable Ultra stable Current Shunts which
ranges from 0.001 to 10£[. The Shunt T/C is less than 10ppm, and
the internal current meter will measures from 1mA to 250Amps. |
|
Shunts Specifications
| Range | Shunt Value | DC Accuracy |
AC accuracy ¡Ø400Hz |
Max input DC / AC rms |
| 200A | 0.001£[ | 0.02% | 0.1% | 250A |
| 20A | 0.01£[ | 0.01% | 0.1% | 30A |
| 2A | 0.1£[ | 0.01% | 0.1% | 4A |
| 200mA | 1£[ | 0.01% | 0.1% | 400mA |
| 20mA | 10£[ | 0.01% | 0.1% | 40mA |
All
shunt types are 4 terminal networks with calibration adjustments for each
network.
AC
accuracy is limited to 100A.
¡@
4 1/2 Digit Ampere Meter
| Range |
resolution (Auto-range) |
DC Accuracy |
AC Accuracy* 50-400Hz |
| 200A | 0.01 / 0.1A |
0.05% of Reading + 2 Counts ¡@ |
0.5% of Reading + 20 Counts
|
| 20A | 0.001 / 0.01A | ||
| 2A | 0.1 / 1mA | ||
| 200mA | 0.01 / 0.1mA | ||
| 20mA | 0.001 / 0.01mA |
For Sine wave input ¡Ù 1800 count.
| General | Accessories |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Options | |
|
|
|
Products list | brand | Model No. | Home