You can select the optimal power supply from a broad range of options, from single-phase 1.5 kVA to a maximum of three-phase 36 kVA.
|
- |
The lineup includes single-phase, single-phase 3-wire, three-phase and multi-phase ( single-phase, single-phase 3-wire, or three-phase selectable) models. |
- |
A multi-phase system (single-phase 3-wire/three-phase) can be configured by connecting multiple units of the same single-phase model. |
*1 Max. 24 kVA (Single-phase model ×2)
*2 Max. 36 kVA (Single-phase model ×3)
*3 1-phase/1-phase 3-wire/3-phase selectable
|
|
The DP series achieves stable output with low distortion, and operates stably with high-capacity capacitor loads. With a variety of output modes and a wide output range.
|
|
- |
AC/DC modes: AC, AC + DC, DC |
- |
Output voltage/frequency |
|
|
|
100 V range |
200 V range |
Resolution |
AC |
Output voltage |
0 V to 155V |
0 V to 310V |
0.1 V |
Frequency |
AC: 40Hz to 550Hz
AC+DC: 1Hz to 550Hz |
0.1 Hz |
DC |
Output voltage |
-220 V to +220V |
-440 V to +440V |
0.1 V |
|
|
|
- |
Output voltage fluctuation: within ±0.15 V (75 V to150 V)/within ±0.30 V (150 V to 300 V)
(DC, 45 Hz to 65 Hz if output current is varied from 0% to 100% of maximum current) |
- |
Maximum peak current: four times the maximum current (RMS value) (corresponds to a capacitor input rectified load with a crest factor of 4) |
- |
Waveform distortion rate: 0.5% max. |
|
In addition to voltage, current and power, the DP series optionally supports measurement of load power factor, crest factor, and up to 40th-order harmonic current. In addition, the series supports measures to control CO2 by displaying CO2 emissions during operation. |
Measurement Items |
- |
Voltage: RMS value, average DC value, peak value |
|
- |
Load power factor* |
- |
Current: RMS value, average DC value, peak value, peak hold value |
|
- |
Crest factor* |
- |
Power: active power, apparent power, reactive power* |
|
- |
Sync frequency |
|
Harmonic current*: up to 40th order |
|
|
CO2 emissions |
* option |
|
Panel operation is a breeze using the large, easy-to-read display.
The design stresses improved ease-of-use and work efficiency for users, with features such as a function for selecting three decided measurement items and displaying them enlarged, and remote operation using a remote controller. |
|
|
|